Zhangjiagang City, Jiangsu Province, China.
+86-512-58810663
These gas sterilizer systems use 100% Ethylene Oxide (EtO) to safely sterilize delicate items that heat and moisture can damage. EtO sterilizers operate at room temperature and atmospheric pressure—safe for electronics, fiber optics, cameras, and plastic or rubber items such as tubing. Sterilizers feature a combination of ventilation and purge systems ensuring that EtO exposure levels are below OSHA prescribed limits. The advanced purge system allows for items to be aerated in the sterilization cabinet for two hours at the end of a cycle and will continue indefinitely until items are removed.
Negative pressure in whole cycle
Chamber pressure is always negative in whole cycle ensures EO gas no leakage to outside during working.
Safety EO gas puncture device
EO gas can automatically punctured by special device in the chamber, no need human intervention.
Chamber pressure real-time monitoring function
Any pressure abnormal in the chamber will cause the machine alarm.
Efficient ventilation system
Automatic pulse Vacuum ventilation ensures EO gas exhaust thoroughly.
Door interlock
The door cannot be opened while program is running.
EO gas concentration monitor(optional)
Accuracy:0.1ppm, portable design.
Accurate parameter control
Sterilization temperature, humidity, pressure and time data are collected by relative sensor and transferred to PLC.
Perfect chamber temperature conducting
Chamber is made of aluminum alloy which has good thermal conductivity makes chamber temperature conducting fast and uniformity.
Dual sterilization temperature choices
Humidity sensor in high accuracy ensures chamber humidity keep in stable level.
Power off memory function
Once power off, the sterilization data will be saved automatically and will automatically continue the sterilization when power on.
Technical parameters:
Leave a Reply