• Anesthesia Machine For Sale, Anesthesia Machine, Electric Anesthesia, Operating Room Anesthesia, Medical Anesthesia, Anesthesia system

S6100 PLUS Medical Instrument Operating Room Electric Anesthesia Machine For Sale

  • Product Description
  • Technical Parameters
  • Video

Anesthesia Machine For Sale, Anesthesia Machine, Electric Anesthesia, Operating Room Anesthesia, Medical Anesthesia, Anesthesia system

Anesthesia system is consist of anesthesia gas delivery system (included anesthesia gas delivery pipe, flow meter, oxygen fault alarm and protection devices), respiratory system, SF-I AGSS system (optional devices), vaporizer (with temperature, pressure and flow compensation function, could equip SE6A or SE6B model, could choose different anesthetics such as Halothane, Enflurane, Isoflurane, Sevoflurane ), anesthesia ventilator, SJ-I anesthesia gas monitor (optional devices) and frame.

Intended Application

S6100 plus anesthesia system?(advanced edition) used?for patients whom need?inhalation anesthesia, controlled breathing, or help to breathe as well as monitoring and display of the patient’s ventilation parameters?in medical department. Suitable for adult and children above 3 years old.

Adaptation disease

Various patients who need anesthetization.

Specifications

Adopts the electric control gas driven mode.

With a variety of alarm, presentation and protection function

With backup power supply.When the power failure is automatically switched to backup power

Use 10.4 “color LCD touch screen shows ventilation parameters, alarm information, and the graphical display.

For adults and children

With volume control, pressure limite varity kinds of working method

Using temperature and flow compensation and zero self-locking anesthetics vaporizer?

O2, N2O flowmeter has the linkage?ensure O2 concentration?25%?

With O2 drive insufficient pressure alarm and N2O truncation protection

Intelligent double system, make the clinical usage more safe

Castors adopt one-key lock function, convenient for clinical usage

Adopt electronic flowmeter

Plug-in type design, could hang other equipment such as monitor, EMR and etc

AGSS and anesthesia gas monitor as optional devices

Working environment

Air source ?? ?? ? O2?N2O?AIR

Air pressure ?????????????????280 kPa ? 600 kPa?

Power supply ? ? 100-240V?

Frequency ? ? ?50 /60HZ

Input power ? ? ?80 VA

Ventilation mode

IPPV

A/C

PCV

SIMV

SIGH

Main technical parameters

Adjustable parameters

Flow gauge

O2????????? ? ?0.1?10 L/min

N2O ????? ? ?0.1?10 L/min

AIR ????? ? ???????0.1?10 L/min

O2?and N2O have internal linkage devices, ensure O2 concentration 25%

Oxygen flush ????????????????35 L/min? 75 L/min

Tidal volume(Vt) ?????????????0, 20 mL ? 1500 mL

Including?

For adult ???250 mL ? 1500 mL

For children ??0 mL? 300 mL

MV ? ? ??????????????For adult?18 L/min

For children ?10 L/min

Frequence?Freq? ???????1 bpm ? 100 bpm

(SIMV mode?1 bpm? 40 bpm ?Except SIMV mode ?Adult 4 bpm? 40bpm, children 20bpm ~ 100bpm)

I/E ????????????????????????Adult 4:1(1:0.25)?1:8

Children 2:1?1:0.5? ? 1:8

(Inspiratory time: 0.2s – 12s)

PEEP???????????? ???? ???0 cmH2O ? 30 cmH2O

Pressure triggering sensitivity?PTr??20 cmH2O ? 0 cmH2O?Based on PEEP?

Flow trigger sensitivity ?FTr? closed? 0.5 L/min ? 30 L/min

Pressure control (PC) ??????????5 cmH2O ? 60 cmH2O

Breathless time (inspiratory breathless, breath at the end of the platform)?TIP?0 ? 6s(?50 % Inspiratory time)

SIGH????????????????????????0(off) ?1/100 ? 5/100

Apnea Ventilation???????????????OFF?5 s ? 60 s

Pressure Limit ??20 cmH2O ? 100 cmH2O

Pressure Rise Time ??????????????0.1 – 2.0s

Monitored Parameters

Frequence?Freq? ? ?????????0 /min ? 100 /min

Tidal volume(Vt) ??????????????????0 mL ? 2000 mL

MV????????????? ? ? ?????0 L/min ? 100 L/min

airway pressure?????????????????????0 cmH2O ? 100 cmH2O

Patients with dynamic lung compliance??1mL/cmH2O ? 1000 mL/cmH2O

Concentration of inspired oxygen ?????15 % ? 100 %

EtCO2 ???????????????????????????0mmHg ~ 152mmHg(or 0-20%)

Inspiratory time ???????????????????0s – 12.0 s

Graphical display:

P-T

F-T

P-V loop

V-T loop

EtCO2 waveform

Alarm and protection

The AC power failure alarm

Internal battery backup low voltage alarm

No tidal volume

High (low) ventilation alarm

High (low)Airway pressure alarm

Pressure limit

Continuous pressure alarm

High (low)Oxygen concentration alarm

Suffocation warning

The maximum limited pressure???12.5 kPa

With O2 drive insufficient pressure alarm and N2O truncation protection

O2, N2O flowmeter has the linkage to ensure O2 concentration?25%?When O2 closed ?N2O closes at the same time.

System failure alarm

Low driven gas pressure alarm

Anesthesia gas Adjustable scope % (volume percentage)
Halothane 0 ~ 5
Enflurane 0 ~ 5
Isoflurane 0 ~ 5
Sevoflurane 0 ~ 8

Products List