| Specification |
pacer (Optional) |
| Type |
VVI |
| Pulse Type |
rectangular,constant current |
| Pulse Amplitude |
0 to 180 mA±10% or 5 mA (whichever is greater). |
| Pacing Rate |
Variable from 30 ppm to 180 ppm +1.5% (increments or decrements by a value of 2 ppm) |
| Multi-Function Electrode (MFE)Pads |
multipurpose defibillation/pacing electrodes |
| Pause |
Pacing pulse frequency reduced by a factor of 4 when activated. |
| Refractory Period |
NC(VVI demand do not have this specification) |
| Pulse Width |
(20+ 1.5 ms) |
| Output Protection |
Fully defibrillator protected and isolated |
| Specification |
EtCO2 (Optional) |
| Type of sensor |
By-pass |
| Technical principle |
Non-dispersive infrared gas analysis NDIR |
| Storage condition |
-40 °C to 70°C,<90% RH, non-condensing |
| Operating conditions |
5°C to 50 °C, 10 to 90% RH, non-condensing |
| Ambient pressure |
55-115kPa |
| Power supply |
5V±5% (max ripple 200 mVp-p) |
| TDP |
Typical value 120mA Excursion calibration typical value 280mA |
| Range |
0-19.7% (0-150mmHg,or 0-20kPa) |
| Resolution |
0.1mmHg |
| Accuracy |
0-40mmHg±2mmHg |
| 41 – 70 mmHg±5% of reading |
| 71 – 100 mmHg±8% of reading |
| 101 – 150 mmHg±10% of reading |
| Respiratory rate |
2-150 BPM |
| Respiratory rate measurement accuracy |
1% ±1BPM |
| Specification |
NIBP (Optional) |
| Measurement unti: |
mmHg/kPa |
| Measurement Range: |
Adult:10~270 mmHg/kPa |
| Pediatric: 10~200 mmHg/kPa |
| Neonatal: 10~135 mmHg/kPa |
| Resolution |
1 mmHg |
| Accuracy: |
Maximum mean error:5 mmHg |
| Maximum standard deviation:8 mmHg |
| Type of sensor |
By-pass |
| Technical principle |
Non-dispersive infrared gas analysis NDIR |
| Storage condition |
-40°C to 70°C, <90% RH, non-condensing |
| Operating conditions |
5°C to 50°C, 10 to 90% RH, non-condensing |
| Ambient pressure |
55-115kPa |
| Power supply |
5V±5% (max ripple 200 mVp-p) |
| TDP |
Typical value 120mA Excursion calibration typical value 280mA |
| Range |
0-19.7% (0-150mmHg,or 0-20kPa) |
| Resolution |
0.1mmHg |
| Accuracy |
0-40mmHg±2mmHg |
| 41 – 70 mmHg±5% of reading |
| 71 – 100 mmHg±8% of reading |
| 101 – 150 mmHg±10% of reading |
| Respiratory rate |
2-150 BPM |
| Respiratory rate measurement accuracy |
1%±1BPM |
| Specification |
SPO2 Moudle (Optional) |
| Measurement Range: |
30 ~ 100% ,±2% between 80% ~ 90%,Others +5% |
| Alarm Range |
User set high limit and low limit |
| Alarm Accuracy |
±10 within setting values. |
| Alarm Time Accuracy |
Less than 12 sec. |