Zhangjiagang City, Jiangsu Province, China.
+86-512-58810663
Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized. … A heat shield of thin polyethylene film can be used with a radiant warmer to reduce heat loss by convection and evaporation.
HBHKN-2000 Infant Radiant Wamer Introduction:
Pre-warm mode, Manual mode and Baby mode;
Set temperature and skin temperature can be displayed separately;
Main body is made of Stainless steel;
Double sides phototherapy radiant box;
Acrylic glass bassinet;
APGAR timer;
Timer to count phototherapy time;
Failure alarm indication;
RS232 connector.
Standard configuration:
It is mainly composed of the warmer module, controller, bassinet, phototherapy, and the stand and so on.
Specification:
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 850VA
Mode of control: Pre-warm mode, manual mode and baby mode are controlled by micro-computer
Baby mode temperature control range: 34.5-37.5°C
Skin temperature display range: 5°C -65°C
Accuracy of skin temperature sensor: -0.2°C
Temperature uniformity of mattress -2°C
Dominant wavelength of blue light: 400-550nm
Timing range of blue light: 0-9999hrs 59min
APGAR timer: audible and visible alarm when the device runs to 50″ – 1′, 4′ 50″ – 5′, 9′ 50″ – 10′ ;
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on.
Average value of general bilirubin radiation on the mattress effective range:
250uW/cm2 ( the above phototherapy equipment)
600uW/cm2(the nether phototherapy equipment)
Operating range: 18°-30°
Ambient air movement rate: 0.3m/s
Package: each unit is packed to one carton
Size(Main body ): 94cm*cm*108cm
Gross Weight: 57kg