Zhangjiagang City, Jiangsu Province, China.
+86-512-58810663
A neonate may need to be transferred from one place to another for providing medical care. Any major disturbance during transport can be harmful to the baby. HB2000 is an incubator transporting babies safely and without disturbance.
HB2000 Infant Warmer Transport Incubator Features:
1. Air mode and Baby mode controlled by micro-computer;
2. Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
3. Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately;
4. Failure alarm indication;
5. Double wall Acrylic glass hood with side door so that thebassinet can be pulled out from it;
6. Natural air flow humidity;
7. The height of whole unit can be adjusted;
8. With Oxygen cylinder and Oxygen supply configuration;
9. Observation lamp.
HB2000 Infant Warmer Transport Incubator Standard configuration:
Main body(including the hood, controller, bassinet, internal battery, and observation lamp) oxygen cylinders, oxygen supply system, skin temperature sensor, IV pole, mattress, adjustable stand.
Optional configuration: >37 degree, overrid temperature mode
AC power supply: | AC220V-230V/50HZ |
DC power supply: | DC12V/10A or DA24V/6A |
Input power: | 400VA |
Model of control: | air mode and baby model controlled by micro-computer |
Air temperature control range:
override mode: |
25 ?-37?
37-38? |
Baby temperature control range:
override mode: |
34-37?
37-37.5? |
Skin temperature sensor accuracy: | ?0.3? |
Temperature variability: | ?1? |
Temperature uniformity: | ?1.5?(mattress is on the horizontal position) |
Noise inside hood: | ?155dB(A)(environment noise ?45dB(A)) |
Failure alarm: | over-temperature?alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm and so on |
Operating at temperature equilibrium: | 90min(set temp 36? ambient 15?) |
Environmental temperature?operating range: | 10-30?(set temp should be higher 3?than environmental temp) |
Ambient air movement rate: | ?1.0m/ |