Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized. The heat output of these devices is usually regulated by servocontrol to keep the skin temperature constant at a site on the abdomen where a thermistor probe is attached.
HB101/102 Infant Incubator Warmer Description:
1. Servo Control Mode: air temp;
2. Instant Parameter Displaying: set temp, air temp, and heat power rate;
3. Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air temp override, fan failure, system failure;
4. >37° Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;
5. Hood with dual-slope: 6 hand-ports; left & right cabinets, top-shelf;
6. RS-232 Output Connector (Use for network system ).
7. Export model: install the guardrail instead of top shelf
HB101 | HB102 | |
Mattress to Hood | 450mm | 450mm |
Observation side of hood | Two incline | Two incline |
Operate port | 6 | 6 |
Front access panel | 826*310mm | 826*310mm |
Mattress tray size | 630*360mm | 630*360mm |
Power supply | ≤400VA | ≤400VA |
Temperature controlled by microcomputer | Air temp | Air and skin temp Air temp |
Air temp control range | 20℃-37℃ | 20℃-37℃ |
37℃-39℃ | 37℃-39℃ | |
Skin Temp control range | ———– | 30℃-37℃ |
———— | 37℃-38℃ | |
Skin Temp sensor accuracy | ———— | ≤0.3℃ |
Temperature fluctuation | ≤0.5℃ | ≤0.5℃ |
Mattress temperature uniformity | ≤0.8℃ | ≤0.8℃ |
Temperature rise time | ≤30min | ≤30min |
Noise level within hood environment | ≤55dB | ≤55dB |
Alarms | ||
Power failure alarm | yes | Yes |
Fan failure alarm | yes | Yes |
Sensor failure alarm | yes | Yes |
High air temp alarm | +1.5℃ | +1.5℃ |
Low air tem alarm | -2.5℃ | -2.5℃ |
High skin temp alarm | —- | +1.0℃ |
Low skin tem alarm | —– | -1.0℃ |
Over air tem alarm | ≤38℃, ≤40℃ | ≤38℃, ≤40℃ |
Over skin tem alarm | ≤38℃, ≤39℃ | ≤38℃, ≤39℃ |
System failure alarm | yes | yes |
Alarm silence function | yes | yes |
Operational condition | ||
Environment temperature | +20℃-+30℃ | +20℃-+30℃ |
Environment relative humidity | 5%-99% | 5%-99% |
Environment air velocity of flow | <0.3m/s | <0.3m/s |
Others | ||
RS232 interface | yes | yes |
Railing | yes | yes |
Infant bed tilt adjustment | ±10° step less | ±10° step less |
Max load of mattress tray | 10kg | 10kg |
IV Pole Max load | 2kg | 2kg |